The purpose of this investigation was to study the reliability of spontaneous otoacoustic emission (SOAE) suppression tuning curve (STC) measurements. Two repetitions of an SOAE STC, using a 4-dB suppression criterion, were obtained from 5 subjects who exhibited stable SOAE level (SD
1.5 dB) and SOAE frequency (±5 Hz). Analyses of variance revealed no significant difference between the SOAE STC trials (p > .05). The SOAE frequency was below the SOAE STC tip, which is similar to the relationship between the probe frequency and a simultaneously masked psychophysical tuning curve. The mean slope of the SOAE STC low-frequency segment was 53.7 dB/octave, whereas the mean slope of the SOAE STC high-frequency segment was 124.8 dB/octave. The mean low- to high-frequency slope ratio was 2.4. The mean Q10 value for the 5 subjects was 5.3. The SOAE STC low- and high-frequency slopes and Q10 were similar to psychophysical tuning curve data obtained in simultaneous masking and physiological tuning curve data.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| All ASHA Journals | AJA | AJSLP | JSLHR | LSHSS |